Respiration Calorimeters For Studying The Respiratory Exchange

Produced by Bryan Ness, Josephine Paolucci and the Online Distributed Proofreading Team at https://www.pgdp.net. (This book was produced from scanned images of public domain material from the Google Print project.)

Chapters

11. Chapter 11

Since the ventilating air-current has a confined volume, in which there are constantly changing percentages of carbon dioxide, oxygen, and water-vapor, it is important to note t...

9. Chapter 9

The carbon-dioxide production, water-vapor elimination, and oxygen absorption of the subject during 1 or 2 hour periods are recorded in a general way by the amounts of carbon di...

3. Chapter 3

In the early experimenting, it was found impracticable to use piping of very small size, as otherwise stoppage as a result of sediment could easily occur. The pipe found best ad...

7. Chapter 7

This apparatus is designed much after the principle of the Regnault-Reiset apparatus, in that there is a confined volume of air in which the subject lives and which is purified...

6. Chapter 6

The chair calorimeter was designed for experiments to last not more than 6 to 8 hours, as a person can not remain comfortably seated in a chair much longer than this time. For l...

8. Chapter 8

The size of the openings of the pipes into the cylinder is so adjusted that the movement of the elevator is regular and moderate whether it is being raised or lowered, thus avoi...

10. Chapter 10

On the one hand, then, there is the correction on the meter itself, which correction is +1.4 per cent (see page 75); and on the other hand the correction on the sample for the t...

2. Chapter 2

In the small room shown at the south side of the laboratory is placed a powerful electric fan which draws the air from above the floor of the calorimeter laboratory, draws it ov...

5. Chapter 5

To vary the current passing through the manganin heating coils in the air-spaces next the zinc wall, a series of resistances is installed connected directly with the rheostat R_...

4. Chapter 4

The thermometers are graduated from 0 deg. to 12 deg. C. or from 8 deg. to 20 deg. C. and each degree is divided into fiftieths. Without the use of a lens it is possible to read...

12. Chapter 12

Some time before the man enters the chamber, an electric lamp of from 16 to 24 candle-power (depending upon the size of the subject) is placed inside of the chamber as a substit...

1. Chapter 1

Produced by Bryan Ness, Josephine Paolucci and the Online Distributed Proofreading Team at https://www.pgdp.net. (This book was produced from scanned images of public domain mat...